Inside a Pulse Oximeter

Follow one measurement from light entering a fingertip to a pulse rate on a screen, and see where an oxygen saturation estimate comes from.

Everything here is an educational model, not a clinical device. The signals are synthetic and the models are simplified. The content has not yet been reviewed by clinical or engineering experts.

laboratories open
in development

The measurement

From light to a number

A pulse oximeter shines red and infrared light into tissue and measures how much comes out the other side. It estimates the pulse rate and the fraction of hemoglobin carrying oxygen (SpO₂). The five laboratories follow the stages of that measurement in order, and each one works on the signal produced by the one before it.

    Laboratories

    Five connected laboratories

    Questions this module answers

    What you will be able to explain

    Each question points to an article and to the laboratory where you can try it.

    QuestionReadLaboratory

    Evaluate

    Where a pulse oximeter falls short

    A pulse oximeter is useful and widely used, but its reading is an estimate. Knowing why it can be wrong is part of understanding the device. Each limit below connects to something you saw in a laboratory.

    Six limits of pulse oximetry
    LimitWhy it happensWhat it can do to the reading
    MovementMovement changes at about the same speed as the pulse, so filters cannot separate them (Laboratory 4).Missed or invented beats, and an unreliable pulse rate and SpO₂.
    Low blood flowCold fingers or poor circulation make the pulse small compared with the noise (Laboratories 3 and 5).Estimates that jump around, or no reading at all.
    Skin pigmentationMelanin absorbs light, and the calibration groups may not represent everyone.In one hospital study, readings of 92 to 96% were followed by blood measurements below 88% in 12% of Black patients and 4% of white patients (Sjoding et al., 2020).
    Other forms of hemoglobinTwo wavelengths can separate only two forms of hemoglobin (Laboratory 1).Carbon monoxide can make the reading look normal when oxygen delivery is poor. Methemoglobin pushes readings toward the mid-80s.
    Calibration rangeCalibration studies seldom include very low saturations (Laboratory 5).Less certainty at low saturation, where it matters most.
    Averaging delayThe device averages over several seconds to reduce noise (Laboratory 4).The display can lag behind a rapid change.

    This is a simplified summary for learning. For sources and detail, see the Photoplethysmography article.

    Check your understanding of the whole module

    Approach

    How each module is organized

    01

    Discover

    What the device is for and the problem it solves.

    02

    Deconstruct

    Its parts: sensors, electronics, and software.

    03

    Understand

    The science and engineering that make it work.

    04

    Experiment

    Change variables in a simulation and watch what happens.

    05

    Engineer

    Design choices, limits, and trade-offs.

    06

    Evaluate

    Accuracy, reliability, and what the device cannot tell us.

    Educational use only. The simulations use synthetic data and simplified models. They cannot be used to measure oxygen saturation or to make any medical decision.