The red and infrared pulses are compared to make one number. A rule then turns that number into an oxygen saturation estimate. See why that rule has to come from measurements, not just from theory.
This is a teaching model, not a medical device. The signals are synthetic, and the calibration line is a commonly taught approximation. It is not any manufacturer’s curve and is not clinically validated. Never use it to judge a real person’s health.
A pulse oximeter does not measure oxygen directly. It measures how much the red and infrared pulses change, forms a ratio, and converts that ratio into a saturation estimate using a calibration curve. In this laboratory you change the true saturation, the pulse strength, and the noise, and watch the estimate respond.
How the ratio of ratios, R, is calculated from the two pulse signals
How R is converted into an SpO₂ estimate
Why theory alone gives the wrong answer in real tissue
How noise makes the estimate jump from beat to beat
Why a reading is an estimate with uncertainty
Key ideas
Read this first
Ratio of ratios
For each color, divide the pulsing part (AC) by the steady part (DC). That gives the relative size of the pulse. Then divide the red result by the infrared result. The answer, R, no longer depends on how much blood is in the finger or how bright the LEDs are. It depends on oxygen.
From R to saturation, in theory
Using the absorption values from Laboratory 1, you can calculate the R that ideal light absorption would give for each saturation. Lower oxygen gives a larger R, because red light is absorbed more.
Why theory is not enough
Real tissue scatters light, so the true R is different from the ideal one. LEDs also vary a little from unit to unit, and the light path depends on the finger. So the relationship is measured, not calculated.
How calibration is done
Volunteers breathe controlled mixtures of gas to lower their oxygen level in a safe, supervised way. Blood samples measure the true saturation while the device records R. A curve is fitted through the pairs. For safety, such studies usually do not go very low, so readings at low saturation are less certain.
A straight line
A straight line, SpO₂ ≈ 110 − 25 × R, is a commonly taught approximation. Real manufacturers use their own curves from their own studies. This laboratory uses the textbook line for teaching.
A reading is an estimate
The number depends on the calibration group, the sensor, the noise, and the person. Studies have found that pulse oximeters can read higher than the true value in people with darker skin. Learn more in the module’s evaluation section.
Simulation
Turn two pulses into one number
Lower the true saturation. R rises, and the estimate follows along the calibration line.
Compare the two estimates in the table. The ideal-theory estimate is lower than the truth, because it ignores scattering.
Raise the noise. The small dots spread out, because R changes from beat to beat.
Lower the pulse strength and raise the noise together. The estimate becomes unreliable.
From ratio to saturation
Live measurements
Value
Pulse size is the RMS of the pulsing part divided by the steady part, averaged over 12 beats. The spread is how far the beat-by-beat estimates vary. The ideal-theory estimate uses the Beer–Lambert ratio from the absorption values of Laboratory 1.
How the estimate is calculated
The two pulse sizes are divided to give R, and R is put into the calibration line.
Limitations
What this model leaves out
In this model, scattering simply makes the true R 1.5 times the ideal value. That factor was chosen so the textbook line fits. In a real finger it depends on the sensor, the finger, and the LED wavelengths.
The red and infrared pulses have exactly the same shape. Real ones differ slightly.
Skin pigment, nail polish, and other forms of hemoglobin are not modeled, and all of them can change a real reading.
The calibration line is a textbook approximation, not any manufacturer’s curve.
The range is limited to 70% to 100%, where this approach is usually tested.