Calibration and SpO₂ Estimation

The red and infrared pulses are compared to make one number. A rule then turns that number into an oxygen saturation estimate. See why that rule has to come from measurements, not just from theory.

This is a teaching model, not a medical device. The signals are synthetic, and the calibration line is a commonly taught approximation. It is not any manufacturer’s curve and is not clinically validated. Never use it to judge a real person’s health.

Go to the simulation

Summary

What you will learn

A pulse oximeter does not measure oxygen directly. It measures how much the red and infrared pulses change, forms a ratio, and converts that ratio into a saturation estimate using a calibration curve. In this laboratory you change the true saturation, the pulse strength, and the noise, and watch the estimate respond.

  • How the ratio of ratios, R, is calculated from the two pulse signals
  • How R is converted into an SpO₂ estimate
  • Why theory alone gives the wrong answer in real tissue
  • How noise makes the estimate jump from beat to beat
  • Why a reading is an estimate with uncertainty

Key ideas

Read this first

Ratio of ratios

For each color, divide the pulsing part (AC) by the steady part (DC). That gives the relative size of the pulse. Then divide the red result by the infrared result. The answer, R, no longer depends on how much blood is in the finger or how bright the LEDs are. It depends on oxygen.

From R to saturation, in theory

Using the absorption values from Laboratory 1, you can calculate the R that ideal light absorption would give for each saturation. Lower oxygen gives a larger R, because red light is absorbed more.

Why theory is not enough

Real tissue scatters light, so the true R is different from the ideal one. LEDs also vary a little from unit to unit, and the light path depends on the finger. So the relationship is measured, not calculated.

How calibration is done

Volunteers breathe controlled mixtures of gas to lower their oxygen level in a safe, supervised way. Blood samples measure the true saturation while the device records R. A curve is fitted through the pairs. For safety, such studies usually do not go very low, so readings at low saturation are less certain.

A straight line

A straight line, SpO₂ ≈ 110 − 25 × R, is a commonly taught approximation. Real manufacturers use their own curves from their own studies. This laboratory uses the textbook line for teaching.

A reading is an estimate

The number depends on the calibration group, the sensor, the noise, and the person. Studies have found that pulse oximeters can read higher than the true value in people with darker skin. Learn more in the module’s evaluation section.

Simulation

Turn two pulses into one number

From ratio to saturation

Live measurements

Value

Pulse size is the RMS of the pulsing part divided by the steady part, averaged over 12 beats. The spread is how far the beat-by-beat estimates vary. The ideal-theory estimate uses the Beer–Lambert ratio from the absorption values of Laboratory 1.

How the estimate is calculated

The two pulse sizes are divided to give R, and R is put into the calibration line.

Limitations

What this model leaves out

To see where pulse oximeters fall short in practice, continue to the evaluation section of the module.