Real biomedical signals are never clean. Add noise to a simulated pulse signal and see when the heartbeat stops being easy to find.
The signal and the noise are synthetic. They are generated by simple equations to teach the ideas. The noise sizes are chosen for teaching and are not measured from any device.
The pulse in a PPG signal is small, often only a few percent of the whole signal. Anything else that changes the light at the detector can hide it. In this laboratory you add four kinds of disturbance to a clean pulse, and lower the blood flow to the finger, then watch how a number called the signal-to-noise ratio responds.
What counts as noise in a physiological signal
Four common problems: electronic noise, baseline drift, movement, and light flicker
Why a weak pulse is harder to read than a strong one
What signal-to-noise ratio (SNR) means and how it is calculated
Why movement is usually the hardest problem to remove
Why a signal sampled too slowly can show a false, slower ripple (aliasing)
Key ideas
Read this first
What is noise?
Noise is anything in the measurement that is not the quantity you want. Some noise is random, like the hiss of electronics. Some is structured, like slow drift or a burst of movement. Noise is a property of the whole measurement system, not only of the sensor.
Why is PPG so sensitive to it?
The heartbeat changes the detected light by only a small fraction of its average level. A disturbance of the same size as that change can hide the pulse completely, even though the disturbance is small compared with the signal as a whole.
What is signal-to-noise ratio?
SNR compares the size of the pulse with the size of the noise, in decibels (dB). Every 20 dB is a tenfold difference in size, and about 6 dB is a doubling. An SNR of 0 dB means the noise is as large as the pulse.
Why is movement the hardest?
A pulse repeats about once or twice a second. Hand and finger movement happens at similar speeds, so a filter that keeps the pulse also keeps much of the movement. Slow drift and fast hiss are easier because they sit away from the pulse in speed.
Prevention beats repair
Filtering can reduce some noise, but it cannot recover information that was buried. Good sensor design, a secure fit, and shielding from room light all reduce noise before it reaches the processor. The next laboratory looks at what filters can and cannot do.
Four common problems in PPG signals
Problem
What it looks like
Typical causes
How engineers respond
Electronic noise
A fuzzy, jittery line
Random fluctuations in the detector and amplifier, worse when little light reaches the detector
A stronger light source (at the cost of battery power), a better amplifier, averaging or smoothing
Baseline drift
The whole trace slowly rises and falls
Slow changes in contact pressure, position, and tissue, including the effect of breathing
High-pass filtering, or subtracting a slowly moving average
Movement artifact
Large, irregular swings, sometimes bigger than the pulse
Movement changes the pressure on the finger and the path the light takes
A secure mechanical fit, a motion sensor (accelerometer) to help detect or cancel it, and marking unreliable segments so they are not used
Light flicker
A fast ripple on the trace
Room light reaching the detector. Mains-powered lights flicker at twice the mains frequency
Shielding the sensor, measuring with the LEDs off and subtracting, and analog filtering before conversion to numbers
Simulation
Add noise to the pulse
Add electronic noise: the line fuzzes up, but the beats are still visible. Keep raising it to find where they disappear.
Add baseline drift: the whole trace wanders, but each pulse keeps its size.
Add movement: two bursts appear, larger than the pulse and similar in speed.
Set light flicker to 100%: a fast ripple appears. The light flickers 120 times a second, but sampling at 100 times a second shows it at 20.
Lower blood flow to 20%: the pulse shrinks while the noise stays the same, so the SNR falls.
Optical signal at the photodetector
Measured signalClean signal (no noise)
Live measurements
Size (RMS)
Share of noise
RMS (root mean square) is a measure of how large a signal is as it varies. Shares are approximate, because the noise sources are added together. The quality words are a rough guide for this laboratory, not a standard.
How the signal-to-noise ratio is calculated
The size of the pulse is divided by the size of the noise, and the result is converted to decibels. The numbers below update as you move the sliders.
Limitations
What this model leaves out
The noise follows fixed patterns, so the same slider position always gives the same picture. Real noise is different every time.
Here, movement is simply added to the pulse. In a real finger, movement also changes the size and shape of the pulse itself.
There is no ambient-light subtraction and no filter before sampling, to show what the raw problem looks like. Real devices include both.
The noise sizes are teaching values. They do not describe any particular sensor.