Laboratory 3 of 5 · Inside a Pulse Oximeter

Noise and Signal Quality

Real biomedical signals are never clean. Add noise to a simulated pulse signal and see when the heartbeat stops being easy to find.

The signal and the noise are synthetic. They are generated by simple equations to teach the ideas. The noise sizes are chosen for teaching and are not measured from any device.

Go to the simulation

Summary

What you will learn

The pulse in a PPG signal is small, often only a few percent of the whole signal. Anything else that changes the light at the detector can hide it. In this laboratory you add four kinds of disturbance to a clean pulse, and lower the blood flow to the finger, then watch how a number called the signal-to-noise ratio responds.

  • What counts as noise in a physiological signal
  • Four common problems: electronic noise, baseline drift, movement, and light flicker
  • Why a weak pulse is harder to read than a strong one
  • What signal-to-noise ratio (SNR) means and how it is calculated
  • Why movement is usually the hardest problem to remove
  • Why a signal sampled too slowly can show a false, slower ripple (aliasing)

Key ideas

Read this first

What is noise?

Noise is anything in the measurement that is not the quantity you want. Some noise is random, like the hiss of electronics. Some is structured, like slow drift or a burst of movement. Noise is a property of the whole measurement system, not only of the sensor.

Why is PPG so sensitive to it?

The heartbeat changes the detected light by only a small fraction of its average level. A disturbance of the same size as that change can hide the pulse completely, even though the disturbance is small compared with the signal as a whole.

What is signal-to-noise ratio?

SNR compares the size of the pulse with the size of the noise, in decibels (dB). Every 20 dB is a tenfold difference in size, and about 6 dB is a doubling. An SNR of 0 dB means the noise is as large as the pulse.

Why is movement the hardest?

A pulse repeats about once or twice a second. Hand and finger movement happens at similar speeds, so a filter that keeps the pulse also keeps much of the movement. Slow drift and fast hiss are easier because they sit away from the pulse in speed.

Prevention beats repair

Filtering can reduce some noise, but it cannot recover information that was buried. Good sensor design, a secure fit, and shielding from room light all reduce noise before it reaches the processor. The next laboratory looks at what filters can and cannot do.

Four common problems in PPG signals
ProblemWhat it looks likeTypical causesHow engineers respond
Electronic noiseA fuzzy, jittery lineRandom fluctuations in the detector and amplifier, worse when little light reaches the detectorA stronger light source (at the cost of battery power), a better amplifier, averaging or smoothing
Baseline driftThe whole trace slowly rises and fallsSlow changes in contact pressure, position, and tissue, including the effect of breathingHigh-pass filtering, or subtracting a slowly moving average
Movement artifactLarge, irregular swings, sometimes bigger than the pulseMovement changes the pressure on the finger and the path the light takesA secure mechanical fit, a motion sensor (accelerometer) to help detect or cancel it, and marking unreliable segments so they are not used
Light flickerA fast ripple on the traceRoom light reaching the detector. Mains-powered lights flicker at twice the mains frequencyShielding the sensor, measuring with the LEDs off and subtracting, and analog filtering before conversion to numbers

Simulation

Add noise to the pulse

Optical signal at the photodetector

Measured signalClean signal (no noise)

Live measurements

Size (RMS)Share of noise

RMS (root mean square) is a measure of how large a signal is as it varies. Shares are approximate, because the noise sources are added together. The quality words are a rough guide for this laboratory, not a standard.

How the signal-to-noise ratio is calculated

The size of the pulse is divided by the size of the noise, and the result is converted to decibels. The numbers below update as you move the sliders.

Limitations

What this model leaves out

For the ideas behind sampling, aliasing, and signal-to-noise ratio in more depth, read Signal processing in biomedical devices. To see how the pulse is generated in the first place, see the Photoplethysmography article.