Educational simulation

PPG Signal Explorer

Change the heart rate, pulse size, and baseline of a simulated pulse oximeter signal, and watch how each one reshapes the light the sensor detects.

This signal is synthetic. It is generated by a simplified equation to teach the ideas, not recorded from a person. Real signals are noisier and more varied.

Go to the simulation

Summary

What you will learn

Photoplethysmography (PPG) is the optical technique inside a pulse oximeter. A light source shines through the fingertip, a photodetector measures how much light arrives, and each heartbeat changes the amount of blood in the way, and so the light. In the simulation below you generate a synthetic PPG signal and change it to see what each setting does.

  • What photoplethysmography is
  • How changes in blood volume change the light the detector receives
  • The difference between the AC and DC parts of the signal
  • How heart rate sets the frequency of the waveform
  • How amplitude and baseline change the look of the waveform
  • Why a PPG waveform is not a measurement of the heart’s electrical activity

Key ideas

Read this first

Anatomy of a PPG waveform
Two beats of the synthetic signal. A real signal is noisier and varies between people.

What is PPG?

Photoplethysmography is an optical technique. Light is sent into tissue, and a photodetector measures how much comes back or passes through. Because blood absorbs light, changes in the amount of blood in the tissue change the detected light.

Why does the waveform repeat?

With each heartbeat the heart pushes blood into the arteries, and the volume of arterial blood in the fingertip rises and then falls as it drains. That cycle repeats once per beat, so the signal repeats with period T = 60 ÷ heart rate. A faster heart rate means a shorter period.

What do AC and DC mean?

DC is the steady part of the signal. It depends on the tissue, the blood that is always present, the sensor’s position, and the light source. AC is the small part that changes with each beat, mainly because of arterial blood volume. The names come from electronics: a steady level and a varying one.

Why does amplitude change?

On a real device, the pulse size depends on perfusion (how much blood is flowing to the finger), sensor placement, contact pressure, skin and tissue properties, and motion. The size of the pulse alone does not tell you blood oxygen saturation. Oximeters compare how light of two colors is absorbed.

PPG is not ECG

A PPG signal is an optical measurement of blood volume at the sensor. An ECG measures electrical activity associated with the heart through electrodes. The two are related, because the heart’s electrical activity triggers the beat, but they measure different things and look different.

Simulation

Change the signal

Optical signal at the photodetector

Live measurements

SelectedCalculated

AC and DC, separated

The detector sees one trace. It is easier to understand as a steady level (DC) plus a small repeating change (AC). This is a simplified model.